USING THE AC/DC ADAPTER
This monitor is designed for operation with batteries or an AC/DC adapter.
Please use only a compatible AC/DC adapter with required voltage and current as indi-
cated in this manual.
Note!
• No batteries are needed when operating with an AC/DC adapter.
• Please unload the batteries when operating with an AC/DC adapter for an extend-
ed period of time.
• Leaving the batteries in the compartment for a long time may cause leakage, which
may lead to damage of the unit.
• Recommend Adapter specification, do not use otherwise:
Model: SINPRO, HPU15-102 (optional)
Rating:
Input: 100 ~ 240V, AC, 47 ~ 63 Hz, 0.4 ~ 0.2 A
Output: 5.99V, DC, 2A,
Note!
When you use the blood pressure monitor with AC/DC adapter, do not position the device
to make it difficult to disconnect the adapter plug.
APPLYING THE CUFF
• Press your brachial artery approximately 1 inch (2 ~ 3 cm) above the elbow on the inside
of your left arm to determine where your strongest pulse is.
• Slide the end of arm cuff furthest from the tube through the metal ring to a loop.
The smooth cloth should be on the inside of the cuff.
• Wrap the cuff on a bare arm or over thin clothing. Thick clothing or a rolled up sleeve will
cause inaccurate blood pressure measurements.
• If the cuff is located correctly, the velcro will be on the outside of the cuff and metal ring
will not touch your skin.
• Put left arm through the cuff loop.
The bottom of the cuff should be approximately
1 inch (2 ~ 3 cm) above the inner elbow.
The tube should lie over the brachial artery on
the inner part of the arm.
• Pull the cuff so that the top and bottom edges
are tightened around your arm.
• When the cuff is positioned properly, press the
velcro firmly against the pile side of the cuff.
• Sit on a chair and lay your forearm on the ta-
ble so that the cuff is at the same level as your
heart.
ENGLISH
30