Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and
should only be used in such environments:
IMMUNITY
IEC 60601
TEST
TEST LEVEL
3V rms
At 0.15-80 MHz
Conducted RF
IEC 61000-4-6
6V rms
At ISM & Radio
Amateur Freq.
Radiated RF
110 V/m at
IEC 61000-4-3
80-2700 MHz
(Proximity
AM Modulation
fields from RF
And 9-28V/m
wireless com-
at 385-6000MHz,
munications
Pulse Mode and
equipment
other Modulation.
IEC 61000-4-3)
The system shall
be tested as
specified in
IEC60601-1-2
Table 9 for proximity
fields from RF
wireless communi-
cations equipment
using the test
methods specified
in IEC 61000-4-3
NOTA 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the device is used exceeds the applicable RF compliance level above, the device
should be observed to verify normal operation. If abnormal performance is observed, additional meas-
ures may be necessary, such as reorienting or relocating the device.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
ENGLISH
COMPLIANCE
LEVEL
3V rms
At 0.15-80 MHz
6V rms
At ISM & Radio
Amateur Freq.
10 V/m at
80-2700 MHz
AM Modulation
And 9-28V/m
at 385-6000MHz,
Pulse Mode and
other Modulation.
The system shall
be tested as
specified in
IEC60601-1-2
Table 9 for proximity
fields from RF
wireless communi-
cations equipment
using the test
methods specified
in IEC 61000-4-3
ELECTROMAGNETIC ENVIRON-
MENT GUIDANCE
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the device,
including cables, than the recom-
mended separation distance calculat-
ed from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Considering to reduce the minimum
separation distance, based on RISK
MANAGEMENT, and using higher IM-
MUNITY TEST LEVELS that are appro-
priate for the reduced minimum sepa-
ration distance. Minimum separation
distances for higher IMMUNITY TEST
LEVELS shall be calculated using the
following equation:
√P
E = 6/d
where P is the maximum power in W, d
is the minimum separation distance in
m, and E is the IMMUNITY TEST LEV-
ELS in V/m.
Field strengths from fixed RF transmit-
ters, as determined by an electromag-
netic site survey,a should be less than
the compliance level in each frequency
range.
b
Interference may occur
in the vicinity of equip-
ment marked with the
following symbol:
38