9. Accuracy: Pressure: ±3 mmHg
10. Wireless communication: Bluetooth V4.0 BLE Only Mode Frequency Band: 2.402--2.480GHz
The Max of RF Output Power :Bluetooth Low Energy -0.5dBm
11. Environmental temperature for operation: 10°C - 40°C (50°F - 104°F)
12. Environmental humidity for operation: ≤85% RH
13. Environmental temperature for storage and transport:-20°C - 55°C (-4°F - 131°F)
14. Environmental humidity for storage and transport: ≤85% RH
15. Environmental pressure: 80kPa - 105kPa
16. Battery life: more than 500 measurements on a full charge, the battery can maintain the
performance characteristics for a minimum of 300 charge cycles
17. The blood pressure measurement system includes accessories, pump, valve, cuff, and sensor.
Note: These specifications are subject to change without notice.
GENERAL SAFETY AND PRECAUTIONS
1. Read all of the information in the User Guide and other provided instructions before operating the unit.
2. Consult your physician for any of the following situations:
a) The application of the cuff over a wound or inflamed area.
b) The application of the cuff on any limb with intravascular access or therapy, or an
arteriovenous (A-V) shunt.
c) The application of the cuff on the arm on the side of a mastectomy.
d) Simultaneous use with other medical monitoring equipment on the same limb.
e) The blood circulation of the user needs to be checked.
3. Do not use this product in a moving vehicle as this may result in inaccurate measurements.
4. Blood pressure measurements determined by this product are equivalent to those obtained by
professional healthcare practitioners using the cuff/stethoscope auscultation method within the
limits prescribed by the American National Standard, Electronic or Automated Sphygmomanometer.
5. If you are using a smart phone to operate the device and a phone call comes in during the
measurement, the measurement process will be terminated automatically. It is thus recommended
that the phone be set in Airplane mode during measurement to avoid interrupting the measurement.
6. If an Irregular Heartbeat (IHB) is detected during the measurement procedure, the IHB symbol will
be displayed. Under this condition, the Smart Blood Pressure Monitor can keep functioning, but
the results may be inaccurate. Please consult your physician for accurate assessment.
The IHB symbol will be displayed under 2 sets of circumstances:
a)The coefficient of variation (CV) of pulse period >25%.
b)The difference of adjacent pulse period is ≥0.14s and more than 53 percent of the total number
of pulses readings falls within this definition.
7. Please do not use any cuff other than that supplied by the manufacturer as this may result in
inaccurate measurements.
8. For information regarding potential electromagnetic or other interference between the Wireless
Blood Pressure Monitor and other devices together with advice regarding avoidance of such
interference, please see ELECTROMAGNETIC COMPATIBILITY INFORMATION. It is suggested
that the Wireless Blood Pressure Monitor should be operated at least 10 meters away from electric
or wireless devices (e.g. routers, microwave oven, etc.)
9. If the blood pressure measurement (systolic or diastolic) is outside the rated range specified in
part SPECIFICATIONS, the app will immediately display a technical alarm on the screen. In this
case, repeat the measurement ensuring that the proper measurement procedures are followed
and/or consult with your medical professional. The technical alarm is preset in the factory and
Pulse rate: ±5%