Descargar Imprimir esta página

Breg VPULSE Manual Del Usuario página 7

Ocultar thumbs Ver también para VPULSE:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 26
Operating Instructions
Using the Start/Stop button to start, change, pause or stop therapies.
Press the button of the desired therapy or combination of therapies. The buttons selected will be
illuminated.
Cold Only
Dynamic Compression Only
Push the Start/Stop button to begin.
To add therapy, first press Start/Stop button, select the therapy button and press the Start/Stop
button again to continue.
To Pause or Discontinue Use.
Press the Start/Stop
Disconnect by
button to stop.
depressing the
side buttons of the
connectors.
12 of 23
Sequential Compression (Calf)
Pad Only
Start/Stop
Remove the Sequential
Compression Therapy pads and
Thermal-Compression pad.
Alarms
When an alarm occurs, the system will stop the therapy session, sound an audible alert and illuminate
in RED the function key of the alarm area.
To stop an alarm, press the function key that is flashing.
Cold Therapy may alarm if:
• Too low or too high a
therapy temperature may be
detected by the system.
Dynamic Compression may
alarm if:
• The time to inflate the pad is
too short or too long.
Sequential Compression may
alarm if:
• The time to inflate the pad is
too short or too long.
Should the system fail to restart after an alarm has sounded and the stated resolutions have not reset
the System, contact Customer Care at Breg, 1.800.321.0607 or 1.760.795.5440 for further assistance.
To resolve:
• Check the tube connection at the
container and at the cold-compression pad.
• Ensure appropriate ice or water level.
• Activate the cold therapy and press start. If
the issue is not resolved, press the cold
therapy button to deactivate.
• If prescribed, ensure the sequential therapy is
activated and press Start/Stop button to
resume sequential therapy. Contact Customer
Care for further instructions.
To resolve:
• Check the tube connection at the container
and at the cold-compression pad.
• Make sure the cold-compression therapy
pads are properly applied to the body and
connected properly.
• Press dynamic compression and then
press the Start/Stop button. If the issue
is not resolved, contact Customer Care for
further instructions.
To resolve:
• Make sure the sequential compression
therapy pads are properly applied to the
calf and connected properly.
• Press sequential compression and then
press the Start/Stop button. If the issue
is not resolved, contact Customer Care for
further instructions.
13 of 23