Using the IPM-400 to Read (cont'd)
Read Pressure
Display: PSI, Pa, Kpa, Mpa, Bar, mBar, Atm, Kgf/cm
inH
O, mmH
O, cmH
2
2
2
This function measures pressure or vacuum, using an optional pressure module adapter and
one of many optional pressure or vacuum modules.
• Relieve all system pressure prior to making any connections or disconnections.
• All pressure connections and test procedures must be performed by a qualified
service person, according to standard engineering practices.
Failure to observe these warnings can result in severe injury or death or equipment
damage.
Over pressurizing the pressure or vacuum module can damage the sensor.
1.
Turn the Selector knob to PRESSURE.
2.
Select the proper module for the pressure range of the process.
3.
Place the pressure or vacuum module in the pressure module adapter.
4.
Plug the pressure module adapter cable into the pressure module adapter jack on the side
of the meter. The meter displays "– – – – – –" for up to 20 seconds while the module
powers up and is verified. The meter displays the range of the module for 3 seconds.
Press the °C/°F/ZERO button to zero the pressure reading.
5.
6.
Carefully connect the module to the pressure point to be measured.
34
All manuals and user guides at all-guides.com
O
Read and understand the instruction manual for pressure
module accessories (IM 1495) before operating or servicing
this equipment. Failure to understand how to safely operate
these tools can result in an accident causing serious injury or
death.
2
, torr, mmHg, inHg,