Using the IPM-400 to Source and Read (cont'd)
Calibrate Current to Pressure (I/P) Transducers (cont'd)
9. Connect the red lead of the meter to the plus (+) input of the transducer and the black
lead to the minus (–).
10. With the toggle switch at SET, adjust the output to the desired value. Calibrate the output
of the transducer in the usual manner.
11. Flip the toggle switch to HI or LO to use the stored values. See the Storing Output Values
section to change the stored values.