Technical data, options, optional equipment
26
The MT-SICS commands listed below is a selected list of available commands. For additional commands and further information
please refer to the Reference Manual "MT-SICS for B-S/L/L-S balances 11780447" downloadable from the Internet under
www.mt.com/pl or www.mt.com/al.
S – Send stable weight value
Command
S
SI – Send value immediately
Command
SI
SIR – Send weight value immediately and repeat
Command
SIR
Z – Zero
Command
Z
@ – Reset
Command
@
SR – Send weight value on weight change (Send and Repeat)
Command
SR
ST – Send stable weight after pressing ± (transfer) key
Command
ST
SU – Send stable weight value with currently displayed unit
Command
SU
Send the current stable net weight value.
Send the current net weight value, irrespective of balance stability.
Send the net weight values repeatedly, irrespective of balance stability.
Zero the balance.
Resets the balance to the condition found after switching on, but without a zero setting
being performed.
Send the current stable weight value and then send continuously the stable weight
value after every weight change.
The weight change must be at least 12.5 % of the last stable weight value,
minimum = 30d.
Inquiry of actual status of the ST function.
As the "S" command, but with the currently displayed unit.