7.2 What if ...?
Error/Error message
íååååì
î____ï
Error 1
Error 2
Error 3
Error 4
0. 0 00 g
Abort
7.3 Connecting L/L-S balances to other METTLER TOLEDO devices
Device
Titrators:
DL31, 36, 38
DL50, 53, 55, 58
MTCom-Bus 310
SQC 14
(statistical quality control)
Spider (industrial scale)
Viper BC (industrial scale)
LC-PVolume (pipette calibration)
LC-PCalc
LC-P45 (application printer)
RS-P42 (report printer)
Cause
Overload
Underload
No stability
• in taring or adjusting (calibration)
• when reference weight for piece counting
is placed on pan
Wrong adjustment weight on pan or none at all
Reference weight (Piece counting, Percent
weighing, Plus-minus weighing) too small
Internal fault
Wrong weighing pan or pan missing or
not empty
Adjustment aborted with the «C» key
No display
• AC adapter not plugged in
• Batteries discharged (only with compact
models)
Connecting cable
RS9–RS9 (m/f)
11101051
RS9–RSopen (m/-)
21900640
RS9–RS9 (m/f)
11101051
RS9–RS9 (m/m)
21252588
RS9–RS9 (m/f)
11101051
RS9–RS9 (m/f)
11101051
RS9–RS9 (m/f)
11101051
RS9–RS9 (m/f)
11101051
Rectification
Remove sample from weighing pan, zero
again (tare).
Check whether weighing pan is
positioned properly.
Wait for stability before pressing key.
Ensure more stable ambient conditions.
Remove weighing pan and clean if
necessary
Place required adjustment weight in
centre of pan.
Increase reference weight.
Contact METTLER TOLEDO customer
service.
Place correct pan or empty pan on
balance.
Check AC power supply.
Plug AC adapter into power supply.
Replace batteries; if using rechargeables
connect instrument to AC power supply.
settings/
Remarks
send continous
AX balances recommended
Appendix