Water Pre-Warming
1.
The ENDO STRATUS
warmer system can maintain the temperature of bottles of sterile water at 37°C (±3°C). However, the
system is not intended to raise the temperature of the water; therefore, pre-warming of the water bottle is
required.
2.
The water bottle must be pre-warmed with a method such as a fluid warmer if warm water is desired.
Water bottle pre-warming should be performed according to the following instructions:
Place sterile water bottle in a fluid warming oven set to 37°±3°C. Do not exceed this temperature.
If warm water is not desired, turn the water warmer on the pump off and place a bottle of room
temperature sterile water on the water bottle holder.
NOTE:
If the water warmer has been enabled and has exceeded the upper temperature set point limit,
the water warmer's switch indicator light will turn blue and the warmer will automatically shut
off. If the switch indicator light is flashing blue, ensure the water warmer cable is plugged into
the back of the unit. If it is connected and the light is flashing, this may indicate a problem with
the unit. Contact technical service at 800-444-4729.
Read and become familiar with all manufacturer's instructions on warming ovens
and water bottles regarding maximum fluid temperatures. Never exceed 40°C
during water pre-warming. Never use a microwave oven to warm a water bottle,
as this could heat the water to a dangerously high temperature or could result in
uneven warming.
CAUTION:
The surface of the water bottle warmer may be hot to the touch.
Operation
1.
After the pre-procedure setup has been completed as outlined in section VIII, flow can be initiated.
2.
Turn the flow control knob to the first position. "L" will illuminate for "LOW". If a higher flow rate is desired,
turn the knob to the next position for "MEDIUM" and one more position for "HIGH". "M" will illuminate
for "MEDIUM" and "H" will illuminate for "HIGH". The knob will turn freely in each direction, and will
increase or decrease the flow rate accordingly. The ability to quickly turn off CO
pressing the flow control knob in. This will immediately turn the flow of CO
flow light indicator. By turning the CO
("L", "M", "H") will restart the flow of CO
3.
If warming of the CO
4.
If the device is attached to a water bottle warmer, and water warming is desired, press the button for water
bottle warmer. The button will illuminate green.
5.
Use the appropriate air/water valve on the GI endoscope and visual feedback of the GI endoscopic system
to manually distend the gastrointestinal tract with CO
Shutdown
1.
When the procedure is over, turn the knob counter clockwise until all lights are off, turn off the water bottle
warmer (if attached) and turn off the CO
2.
Turn off the main power switch and disconnect the tubing from the front of the unit.
3.
Discard the tubing according to the manufacturer's instructions for proper disposal.
INTERNATIONAL_ IFU-52 (CO2 INSUFLATOR PUMP) REV H .indd 13
®
CO
insufflator can be used to control the water bottle warmer system. The
2
flow control knob off and then back to your desired flow rate setting
2
. See section V for flow rates.
2
is desired, press the button for the CO
2
heater.
2
off and extinguish the CO
2
heater. The button will illuminate green.
2
.
2
13
flow is available by
2
2
IFU-52/ REV H
12/11/2015 9:31:57 AM