VI. Front panel controls/connections
NOTE:
Note: All status indicator lights will be green during normal operation.
1. On/Off push button
Turns main power on or off to the unit
2. Water warmer On/Off
Push to turn water warmer on or off
3. CO
Flow control
2
Controls flow rate of CO
and "H" which are illuminated to indicate the flow rate setting. The ability to quickly turn off CO
available by pressing the flow control knob in. This will immediately turn the flow of CO
the CO
flow light indicator. By turning the CO
2
rate setting ("L", "M", "H") will restart the flow of CO
Low – 1.4L/min
Medium – 2.4L/min
High – 3.5L/min
*Flow rates shown are maximum free flow measurements at output of unit
Low gas pressure is indicated by the words "LOW PRESSURE" flashing to notify the user that the gas input
pressure has fallen below 25PSI.
4. CO
gas output connector
2
Connect to tubing and GI endoscope.
Standard male luer lock connector.
5. CO
gas heater On/Off
2
Push to toggle CO
1
3
4
INTERNATIONAL_ IFU-52 (CO2 INSUFLATOR PUMP) REV H .indd 9
output. The three levels are represented by three green indicators of "L", "M",
2
gas heater On or Off
2
2
(The appearance of your device may differ slightly from picture shown above)
flow control knob off and then back to your desired flow
2
. The flow rates are as follows*:
2
5
9
flow is
2
off and extinguish
2
IFU-52/ REV H
12/11/2015 9:31:56 AM