"Flush" function
(Ref. 6 inside cover)
The "flush" function can be used when
the type of powder needs to be changed,
to make sure the powder circuit has been
properly cleaned of the previously used
powder.
NOTE: If both handpieces are stowed
in their housings, the "flush" function
cannot be activated.
5.2.3

Symbols

(Ref. 10 inside cover)
combi touch is provided with a diagnostic
circuit that allows the detection of
malfunctions and viewing of their type
on the touch keyboard by means of a
symbol.
To
help
the
user
malfunctioning component, there are
three symbols described in
on page 112
.
5.3

Safety Requirements Before and During Use

WARNING: Exclusively use original
Mectron inserts, accessories and
replacement parts.
WARNING: Before starting work always
ensure that you have all replacement parts
(handpiece, inserts, keys) to use should there
be downtime due to a fault or problem.
WARNING: Checking the condition of
the device before treatment. Always check
that there is no water under the appliance.
Before every treatment always check the
perfect operation of the appliance and
efficiency of the accessories. If operation
anomalies are observed do not carry out the
treatment. Contact an Authorised Mectron
Service Centre if the faults concern the
device.
identify
the
Chapter 12.1
use: All parts and reusable accessories (new
or returned from an Authorised MECTRON
Service Centre) are delivered in NON-STERILE
conditions and must be treated before each
use, according to the instructions provided in
Chapter 8 on page 49
After each treatment, clean and sterilise all
parts and reusable accessories according to
the instructions in
leave liquids in the tank for long periods of
time. The tank must be filled after having
cleaned and sterilised all parts, and before a
treatment. If the tank has been filled without
having used the device, it must be emptied
at the end of the day, and all parts and
accessories must be cleaned and sterilised.
into contact with the machine core unit or
with the pedal.
WARNING: Infection control. First
. Subsequent uses:
Chapter 8 on page 49
WARNING: Infection control. Do not
WARNING: The patient must not come
USE
refill
refill
.
29