The Ultravision Ionwand must be inserted in the correct orientation. Failure to do so may cause damage to the connector
or socket.
4 – Proximity indicator LEDs
When the generator is switched on, the proximity indicator LEDs illuminate green. In order to function correctly, contact be-
tween the Ionwand and patient tissue or other surgical instruments should be minimised. Both proximity indicator LEDs will
flash when the Ionwand comes within approximately 2mm (1/12") of patient tissue. There is also an audible indicator that will
sound. When the contact is removed the LEDs will stop flashing and the audible alarm will stop sounding.
Warning
Do not allow the Ionwand to come in to contact with the uninsulated portion of an electrosurgical instrument during operation.
Continued contact for in excess of 10 minutes may result in an electrical discharge from the electrosurgical instrument that
may be sufficient to cause atrial fibrillation depending on proximity to the thoracic cavity.
5 – Battery level indicator
With a new, fully-charged battery, the battery level indicator displays the approximate power remaining as follows:
The red flashing LED is also accompanied by an audible indicator.
Where the Ultravision Mains Converter is used to power the Ultravision system, the Green LED is permanently illuminated
on this indicator when the unit is switched on. If the Yellow or Red LED is illuminated, then the Mains Converter should be
replaced.
Caution
The battery should not be used for a surgical procedure when the red flashing LED is present and should be re-charged.
Failure to do so may result in loss of power during the procedure, leading to impaired visibility.
Be Aware
The battery may lose energy during transportation, storage and delivery particularly with prolonged exposure to higher ambi-
ent temperatures. It is recommended that the battery be fully charged before first use of the Ultravision system.
It is recommended that the battery be replaced once it is no longer able to deliver 6 hours of power when fully charged. Con-
tact Alesi Surgical or its local distributor to purchase replacements.
6 – Generator fault indicator LED
This LED is not illuminated during use. If there is a problem within the generator, the LED will illuminate red and flash. This is
also accompanied by an audible indicator.
Warning
Do not attempt to use the Ultravision generator when the generator fault indicator is lit. Contact Alesi Surgical or its local
distributor to arrange for repair or replacement of the generator unit.
Ultravision™ generator – Side panel view
1 – Battery/Mains Converter insertion mechanism
1 – Battery/Mains Converter insertion slot and release mechanism
The battery or Mains Converter is inserted into the slot on the right-hand side of the Ultravision generator. It is subsequently
12
Green LED
Amber LED
Red LED
Red flashing LED
>8hr power remaining
2-8hr power remaining
<2hr power remaining
<30min power remaining