Chapter 9: Electromagnetic guidance

Electromagnetic emissions – Ultravision Generator
With reference to the 4
Edition of IEC 60601-1-2:2014 The Ultravision generator is intended for use in the Special Environ-
th
ment of high powered medical electrical equipment associated with powered surgical devices. The customer or the user of
the Ultravision generator should assure that it is only used in such an environment.
Emissions test Compliance Electromagnetic environ-
ment
– guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions IEC 61000-3-3
Electromagnetic immunity – Ultravision Generator
The Ultravision generator is intended for use in the electromagnetic environment specified below. The customer or the user
of the Ultravision generator should assure that it is used in such an environment.
I
test
MMUNITY
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field IEC 61000-
4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
30
IEC 60601-1-2 test level
± 8 kV contact
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
100kHz repetition frequency
±
1 kV Differential Mode
±
2 kV Common Mode
0 % U
; 250/300 cycle
T
30 A/m
Compliance
Electromagnetic environment - guidance
Group 1
The Ultravision generator uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interfer-
ence in nearby electronic equipment. RF emissions
Class A
The Ultravision generator is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
Class A
supply network that supplies buildings used for
domestic purposes.
Complies
Compliance level
± 8 kV contact
± 15 kV air
± 2 kV for power supply
lines
± 1 kV for input/output lines
±
1 kV Differential Mode
±
2 kV Common Mode
0 % U
; 250/300 cycle
T
30 A/m
Electromagnetic environment
guidance
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the electrosurgical
Electrosurgical generator requires
continued operation during mains power
interruptions, it is recommended that
the Ultravision generator be powered
from the identified battery accessory.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment.