Descargar Imprimir esta página

Keysight EDU33210 Serie Manual De Instrucciones página 96

Publicidad

Idiomas disponibles

Idiomas disponibles

Front Panel Operations
1. Select Pulse waveform: Press [Waveform] > Pulse.
2. Select period instead of frequency: Press [Units] >
3. Set the period: Press [Parameter] > Period. Use the numeric keypad or the knob and arrows to set a desired value. If
you use the keypad, select a unit prefix to finish.
SCPI Command
[SOURce[1|2]:]FUNCtion:PULSe:PERiod {<seconds>|MINimum|MAXimum|DEFault}
Pulse Width
Pulse width is the time from the 50% threshold of a pulse's rising edge to the 50% threshold of the next falling edge.
– Pulse width: up to 1,000,000 s (see restrictions below). The default pulse width is 100 μs. The minimum pulse
width is 16 ns.
– The specified pulse width must also be less than the difference between the period and the minimum pulse width.
– The instrument will adjust the pulse width to accommodate the specified period.
Front Panel Operations
Press [Waveform] > Pulse > Pulse Width. Use the numeric keypad or the knob and arrows to set a desired value. If
you use the keypad, select a unit prefix to finish.
96
Frequency
Periodic > Frequency Periodic.
Keysight EDU33210 Series User's Guide

Publicidad

loading