Descargar Imprimir esta página

BRAND Transferpette S-8 Instrucciones De Manejo página 55

Ocultar thumbs Ver también para Transferpette S-8:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 114
6 Accuracy table
Volume
Partial
range
volume
[μl]
[μl]
0.5...10
  10
    5
    1
5...50
  50
  25
    5
10...100
100
  50
  10
20...200
200
100
  20
30...300
300
150
  30
*A = Accuracy, CV = Coefficient of Variation
20 °C
!
Ex
Final test values based on the nominal volume (= max. volume) printed on
the device and the specified partial volumes at the same temperature (20
°C/68 °F) of the device, surroundings and distilled water, in accordance
with DIN EN ISO 8655.
Transferpette® S-8/-12 997208
A* ≤ ±
CV* ≤ %
1.6
1.0
2
2
8
6
0.8
0.4
1.4
0.8
6
3
0.8
0.3
1.4
0.6
4
2
0.8
0.3
1.4
0.6
4
1.5
0.6
0.3
1.2
0.6
3
1.5
Sub
Recommended
steps
tip type
[μl]
[μl]
0.01
0.5...20
0.05
2...200
0.1
2...200
0.2
2...200
0.5
5...300
Operating manual
55

Publicidad

loading

Este manual también es adecuado para:

Transferpette s-12