Fluid pressure hazard. Removal of a sensor from a pressurized vessel can be dangerous. Reduce
the process pressure to below 7.25 psi (50 kPa) before removal. If this is not possible, use extreme
caution. Refer to the documentation supplied with the mounting hardware for more information.
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
6.1 Maintenance schedule
Table 1
shows minimum times for regular maintenance tasks. Perform maintenance tasks more
frequently for applications that cause electrode fouling.
Task
Clean the sensor
on page 13
Examine the sensor for damage
Calibrate the sensor
6.2 Clean the sensor
Pre-requisite: Prepare a mild soap solution with a non-abrasive dishwashing detergent that does not
contain lanolin. Lanolin leaves a film on the electrode surface that can degrade the sensor
performance.
Examine the sensor periodically for debris and deposits. Clean the sensor when there is a buildup of
deposits or when performance has degraded.
1. Use a clean, soft cloth to remove loose debris from the end of the sensor. Rinse the sensor with
clean, warm water.
2. Soak the sensor for 2 to 3 minutes in the soap solution.
3. Use a soft bristle brush to scrub the entire measuring end of the sensor.
4. If debris remains, soak the measuring end of the sensor in a dilute acid solution such as < 5%
HCl for a maximum of 5 minutes.
5. Rinse the sensor with water and then return to the soap solution for 2 to 3 minutes.
6. Rinse the sensor with clean water.
Always calibrate the sensor after maintenance procedures are done.
Section 7 Troubleshooting
7.1 Intermittent data
During calibration, data is not sent to the datalog. Thus, the datalog can have areas where the data is
intermittent.
W A R N I N G
W A R N I N G
C A U T I O N
Table 1 Maintenance schedule
90 days
X
X
Set by regulatory agencies or experience
1 year
English 13