Warning
The device is not
calibrated.
Replace a sensor.
Calibration is in
progress.
Outputs on hold
Linear temperature
compensation is out of
range.
Temperature
compensation table is
out of range.
Incorrect user
concentration table.
Incorrect built-in
temperature table.
Incorrect built-in
concentration table.
Section 8 Consumables and accessories
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Consumables
Description
Conductivity standard solution, 100 µS/cm
Conductivity standard solution, 12.88 mS/cm
Conductivity standard solution, 12.88 mS/cm
Conductivity standard solution, 146.9 µS/cm
Conductivity standard solution, 146.9 µS/cm
Conductivity standard solution, 147 µS/cm
Conductivity standard solution, 147 µS/cm
Conductivity standard solution, 180 µS/cm
16 English
Table 4 Warning list (continued)
Description
The sensor has not been calibrated.
The Sensor days counter is more
than the interval selected for sensor
replacement. Refer to
Configure the
sensor
on page 6.
A calibration was started but not
completed.
During calibration, the controller
outputs were set to hold for a
selected time.
The user-defined linear temperature
compensation is out of range.
The user-defined temperature
compensation table is out of range.
The concentration measurement is
outside of the range of the user table.
The measured temperature is outside
of the range of the built-in
temperature compensation table.
The concentration measurement is
outside of the range of the built-in
concentration table.
W A R N I N G
Resolution
Calibrate the sensor.
Replace the sensor. Reset the Sensor days
counter on the Diagnostics/Test > Reset
menu (or Diagnostics/Test > Counter menu.
Return to calibration.
The outputs will become active after the
selected time period.
The value must be between 0 and 4%/°C;
0 to 200 °C.
The temperature is above or below the
temperature range defined by the table.
Make sure that the user table is set for the
correct measurement range.
Make sure that the temperature
compensation is configured correctly.
Make sure that the concentration
measurement is configured for the correct
chemical and range.
Quantity
50 mL
500 mL
20 mL, 20/pkg
50 mL
500 mL
20 mL, 20/pkg
125 mL
100 mL
Item no.
2971826
C20C250
2771320
2974226
2974249
2771320
LZW9701.99
2307542