Descargar Imprimir esta página

Hach 8398 Serie Manual Del Usuario página 5

Ocultar thumbs Ver también para 8398 Serie:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 44
Section 3 Installation
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
3.1 Connect the sensor cable to an SC Controller
1. Install an ultrapure conductivity module in the SC Controller. Refer to the installation instructions
supplied with the ultrapure conductivity module. Refer to
on page 16 for ordering information.
2. Connect the wires of the conductivity sensor cable to the ultrapure conductivity module. Refer to
the wiring instructions supplied with the ultrapure conductivity module.
Note: The ultrapure conductivity module converts the analog signal from the sensor to a digital signal.
3.2 Mounting
Explosion hazard. Make sure that the mounting hardware for the sensor has a temperature and
pressure rating sufficient for the mounting location.
For the best performance, make sure that the Pt100 temperature sensor is fully in the sample and
the sensor opening points in the direction of flow. Refer to
clearance (minimum) around the sensor to prevent measurement interference.
Refer to
Figure 3
for mounting examples.
Figure 2 Minimum clearance
1 Pt100 temperature sensor
C A U T I O N
Consumables and accessories
W A R N I N G
Figure
2 Minimum gap of 1 cm (0.4 inches)
2. Make sure that there is a 1 cm of
English 5

Publicidad

loading

Este manual también es adecuado para:

8332