4.2
Function checks
Check that the accessories do not show any visible
damage.
Switch the Aesculap device to ready status, see
instructions for use of the Aesculap device.
Check that no HF instrument is connected to the
Aesculap device.
Function tests:
– Connect the plug of the double pedal foot
switch at the foot switch connection socket of
the Aesculap device, but do not press the foot
switch. Expected reaction of the device: No
activation.
– Press the double pedal foot switch. Expected
reaction of the device: Activation of HF or JET.
– Release the double pedal foot switch. Expected
reaction of the device: Activation of HF or JET
is stopped.
Note
Device reactions at variance with those listed above
may indicate a fault of the Aesculap device or the dou-
ble pedal foot switch, or both. If possible repeat the test
with another Aesculap device!
4.3
Safe operation
Risk of injury and/or malfunc-
tion!
WARNING
Risk of burns to the patient
due to inadvertent activation
of the Aesculap device!
WARNING
Hazard risks due to inadequate
preparation or faulty Aesculap
device!
WARNING
Press the double pedal foot switch.
The Aesculap device is activated.
Always carry out a func-
tion check prior to using
the product.
Immediately switch off the
Aesculap device at the
mains power OFF switch.
Operate the double pedal
foot switch with appropri-
ate caution.
Ensure that the Aesculap
device is fully operational
without fault.
Check that no conductive
fluids (e.g. blood, amniotic
fluid) have entered the
double pedal foot switch.
Check that there is no
short circuit in the double
pedal foot switch or the
foot switch cable.