The 200-006R is intended for use in the electromagnetic environment specified below. The customer or the user of the 200-006R
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
Test
Test Type
levels
ESU Compatibility
none
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the 200-006R is used exceeds the applicable RF compliance level above, the 200-006R
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the 200-006R.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m
b
IEC60601-1-2 Table 4 RF Electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity
Compliance level
3Vrms
3 V/m
Frequency
Mains
Range
Supply
The 200-006R
has been
tested for
compatibility
with Electrical
Surgical
Equipment
(ESU) used in
a typical
hospital
environment
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the 200-006R,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1,17 √P
80 MHz to 800 MHz
d = 1,17 √P
80 MHz to 800 MHz
d = 2,33 √P
800 MHz to 2,5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
be less than the compliance level in each frequency
range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Standby
Run Mode
Port(s)
Mode
Tested
.
a
, should
Results
PASS/ FAIL