The 200-006R is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the 200-006R can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the 200-006R as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
IEC 60601-1-2 Table 9 Immunity to RF wireless communications equipment
Test frequency
Band
(MHz)
(MHz)
385
380 –390
450
430 – 470
710
745
704 – 787
780
810
870
800 – 960
930
1 720
1 845
1 700 – 1 990
1 970
2 450
2 400 – 2 570
5 240
5 500
5 100 – 5 800
5 785
IEC60601-1-2 Table 6 RF Immunity
Recommended separation distances between
portable and mobile RF communications equipment and the 200-006R
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1,17 √P
0.12
0.37
1.17
3.69
11.67
Service
TETRA 400
GMRS 460, FRS 460
LTE Band 13, 17
GSM 800/900, TETRA 800,
iDEN 820, CDMA 850,
LTE Band 5
GSM 1800; CDMA 1900;
GSM 1900; DECT;
LTE Band 1, 3, 4, 25;
UMTS
Bluetooth, WLAN,
802.11 b/g/n, RFID 2450,
LTE Band 7
WLAN 802.11 a/n
m
80 MHz to 800 MHz
d = 1,17 √P
0.12
0.37
1.17
3.69
11.67
Immunity test level
Modulation
Pulse modulation b)
18 Hz
FM c)
± 5 kHz deviation
1 kHz sine
Pulse modulation b)
217 Hz
Pulse modulation b)
18 Hz
Pulse modulation b)
217 Hz
Pulse modulation b)
217 Hz
Pulse modulation b)
217 Hz
800 MHz to 2,5 GHz
d = 2,33 √P
0.23
0.74
2.33
7.38
23.33
Compliance
(V/m)
Levels
27
7
28
7
9
3
28
10
28
10
28
28
9
9