Verify the assembly of the 006 Hanger Replacement Kit. If parts are loose arrange them as shown.
Attach the replacement hanger onto the UPS.
- Using a Phillips-head screwdriver, tighten the two screws of the replacement hanger assembly to the body of the UPS.
- Check for rotation of the hanger and click into place.
Service or Repair
The UPS is a reusable device with no user-maintainable/serviceable features.
To schedule service or repair, contact Microline Surgical customer service at the number located on the back cover of this
service manual.
WARNING: No modification of this equipment is allowed.
7.0
Troubleshooting
General
Before starting troubleshooting, try to eliminate all possibilities of operational errors that may have caused the problem:
- Check the unit for visible signs of physical damage.
- Verify that all cords are connected and attached properly.
If the malfunction remains, use the troubleshooting guides below. Perform actions step by step and check that the malfunction
is eliminated.
The troubleshooting guides below are focused only on technical problems. Information about clinical related problems can be
found in the device instructions for use.
Correcting Malfunctions
Situation
No UPS output
- Instrument plugged in
- Footswitch indicator flashing yellow
No UPS output
Abnormal UPS output
- Instrument plugged in
- Footswitch plugged in
- Footswitch indicator flashing yellow
8.0
Maintenance
Routine Maintenance
Check the power cord each time you use the UPS or at intervals recommended by your institution. Replace the power cord if
you find exposed wires, cracks, frayed edges or a damaged connector with Microline part number:
- PRT0495 (power cord for use in Continental Europe [CEE 7/7 plug])
- PRT0496 (power cord for use in Italy [CEI 23-16 plug])
- PRT0497 (power cord for use in the United Kingdom [BS 1363 plug])
- PRT0883 (power cord for use in USA, Canada [NEMA 5-15P plug])
- PRT0939 (power cord for use in Switzerland [SEV 1011 plug])
- PRT0940 (power cord for use in Denmark [SRAF 1962/DB 16/87 plug])
- PRT0945 (power cord for use in Australia, New Zealand [AS 3112 plug])
Possible Cause
Footswitch required for instrument operation
Thermal shutdown due to continuous use
Footswitch not required for instrument
operation
Solution
Connect Microline PowerPack
Footswitch to UPS
Allow the UPS to cool down for 10
minutes
Disconnect Microline PowerPack
Footswitch from UPS