Descargar Imprimir esta página

DJO Global Intelect Legend 2 Manual Del Usuario página 20

Publicidad

Idiomas disponibles

Idiomas disponibles

20
SYSTEM
WAVEFORMS (CONTINUED)
IFC Premodulated (Traditional 2 Pole)
Advice on size and type of electrodes to be used is given in
device GUI "treatment guidelines" feature
Premodulated Current is a medium frequency waveform.
Current comes out of one channel (two electrodes). The
current intensity is modulated: it increases and decreases
at a regular frequency (the Amplitude Modulation
Frequency).
Output Mode
Available on Channel
Treatment Time (STIM)
Treatment Time (COMBO)
Mode Selection
Output Intensity
Beat Fixed (Sweep Off)
Cycle Time
Frequency
Sweep Low Beat Frequency
Sweep High Beat Frequency
IRMS
DC component
Russian
Advice on size and type of electrodes to be used is given in
device GUI "treatment guidelines" feature
Russian Current is a sinusoidal waveform, delivered in
bursts or series of pulses.
Output Mode
Available on Channels
Treatment Time
Mode Selection
Output Intensity
Burst Frequency
Carrier Frequency
Cycle Time
Duty Cycle
Ramp
IRMS
INTELECT
LEGEND 2 USER MANUAL
®
Electrodes
1, 2, 3, 4
1-60 Min
1-30 Min
CC or CV
0-100 mA (CC) 0-100 V (CV))
Carrier
1-200 Hz
Continuous or User Defined
2,000-10,000 Hz
1- 199 Hz
2-200 Hz
0-55mA
No
Electrodes
1, 2, 3, 4
1-60 min
CC or CV
0-100 mA (CC) 0-100 V (CV)
1-100 bps
2,500 Hz
Continuous or User Defined
10%, 20%, 30%, 40%, 50%
0-5 sec
0-39mA
DC component
VMS™ Burst
Advice on size and type of electrodes to be used is given in
device GUI "treatment guidelines" feature
VMS Burst is a symmetrical biphasic waveform delivered
in a burst format. Because the pulse is relatively short,
the waveform has a low skin load, making it suitable for
applications requiring high intensities, such as muscle re-
education protocols.
Output Mode
Available on Channels
Treatment Time
Mode Selection
Output Intensity
Anti-
Burst Frequency
Channel Mode
Cycle Time
Duration
Ramp
Set Intensity
Setting in Reciprocal and Co-Contract modes
IRMS
DC component
DC (Direct Current)
Advice on size and type of electrodes to be used is given in
device GUI "treatment guidelines" feature
Galvanic Current is a direct current flowing in one
direction only.
The current can be continuous or interrupted.
Output Mode
Available on Channels
Treatment Time
Mode Selection
Output Intensity
Cycle Time
Polarity Reversal
With Polarity Reversal On, Polarity will change in
the middle of the treatment time.
IRMS
DC component
EN
Electrodes
1, 2, 3, 4
1-60 min
CC or CV
0-140 mA (CC) 0-140 V (CV)
Off or On
1-200 bps
Single, Reciprocal, Co-Contract
Phase
Continuous or User Defined
30-400 μsec
0-5 sec
Individual/both Channel Intensity
0-50mA
Electrodes
1, 2, 3, 4
1-60 min
0-40 mA (CC)
Continuous, or User Defined
On or Off
0-44mA
No
CC
Yes

Publicidad

loading

Este manual también es adecuado para:

12-501012-5011