PATIENT PREPARATION
ULTRASOUND PATIENT PREPARATION
1.
Examine the skin for any wounds and clean the skin
2.
View the Applicator recommendation in the treatment
guidelines.
3.
Review guidelines for Ultrasound (as a reference
point only) on the treatment review screen prior to
administering treatment.
NOTE: Applicators are available in the sizes shown below:
Applicator Preparation and Use
1.
Clean applicator before each therapy session with
warm soapy water, check the applicator has no cracks
prior to use.
2.
Liberally apply transmission gel to the treatment area
on the patient.
3. Move the applicator during therapy session in a
circular motion. The area treated should be:
• Twice the diameter of the applicator
• For 5cm
and 10 cm2 US applicator: three times the
2
diameter of the applicator if output power > 4 W,
Continuous mode.
INTELECT
LEGEND 2 USER MANUAL
®
10 cm
2
5 cm
2
2 cm
2
1 cm
2
The applicator should always be held by the grip and not
by the Ultrasound Applicator head.
4.
If US Coupling is "On", the Applicator is properly
coupled to the patient and administering ultrasound
when the LED is constantly illuminated. If the
applicator head becomes uncoupled the LED on the
head will flash. If "US coupling" setting is ON, several
beeps will be also heard until the head is coupled
again. Treatment time stops during uncoupling.
NOTE: Ultrasound output will continue to be emitted in all
US coupling modes even if the applicator is uncoupled.
The output power is reduced to a very low level to prevent
ultrasound head warming.
For ULTRASOUND OPERATION, refer to page 32