Descargar Imprimir esta página

Aesculap Flow 50 Instrucciones De Manejo página 96

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

General Information
EN
14
WARNING!
Before using the insufflator to endoscopically harvest vessels, please check
whether the instrument used is intended for CO
WARNING!
Pneumoperitoneum
When a vessel is harvested from the leg of a patient with a perforated groin, it is
possible for CO
to reach the abdomen and cause a pneumoperitoneum. Make
2
sure the abdomen does not fill with CO
3.3.5
Warnings Fieldflooding Mode (Cardio Indication)
WARNING!
A gas flow of 10 l/min should not be used longer than up to 30 seconds.
WARNING!
When using accessories not specified in the user manual, make sure the CO
not directly applied to soft tissue focusse.
3.4
General Precautions
CAUTION!
Continuous operation
After 24 hours of continuous operation, a device self-test must be carried out.
• Switch device off and on again.
CAUTION!
Endoscope
The device may only be connected with endoscopes designed for and featuring
the technical specification permitting such a combined use. Any utilized endo-
scopes must comply with the most recent versions of IEC 60601-2-18 and ISO
8600. Combining/connecting with other devices generates a medical electrical
system (MES). The system configurator is responsible for compliance with the
standard IEC 60601-1 / EN 60601-1 in its latest version.
CAUTION!
Electrical Interference
(See Chapter 11 Electromagnetic Compatibility). Care was taken during the de-
velopment and testing of this device that electrical interference from this device
or other devices or instruments was practically eliminated. However, if you still
detect or suspect such interference, please follow these suggestions:
• Move this, the other or both devices to a different location
• Increase distance between used devices
• Consult an electro-medical expert
insufflation.
2
during surgery.
2
is
2

Publicidad

loading

Este manual también es adecuado para:

Pg150