11. Quality Control
11.1 Onboard QC Test
The qLabs
ElectroMeter Plus also measures the clotting time in the
®
Quality Control (QC) zone of PT-INR or APTT strip. If the QC result fails
to fall within a predetermined range, the qLabs
display an error code rather than give a possible erroneous PT-INR or
APTT result.
This safety measure guards the user against situations in which the
qLabs
PT-INR or APTT test strip may have been subjected to very high
®
temperatures or humidity (which could happen if the foil pouch is torn or
punctured).
11.2 Liquid QC Test
Micropoint Biotechnologies Co., Ltd. has optional liquid controls for
qLabs
system. The qLabs
®
with qLabs
PT-INR and APTT test strips to verify the qLabs
®
system performance, respectively. It is recommended that two levels of
liquid controls are used to perform the test.
Please follow the QC test procedures below:
TEST TYPE
PATIENT TEST
QC TEST
OK
CONTROL CODE
PT
9324428
LOT: 3052E0503
OK
PT-INR and APTT liquid controls are used
®
Turn on the meter by inserting a new test
strip and enter the "Test Type" screen. Use
the left and right buttons to select the test
type. Select the "QC control" and then press
"OK" to accept to enter the "Control Code"
screen.
When the lot number input function is
enabled, press " " to edit the control code
and lot number or use "
control code and lot number via barcode
scanner. Go to Section 9.6 for specific
information. Press "OK" to accept and enter
the "Strip Code" screen.
29
ElectroMeter Plus will
®
PT-INR
®
" to enter the
EN