the Partial Thromboplastin Time (PTT). It can provide a more precise and
sensitive assay.
Micropoint Biotechnologies Co., Ltd. provides various combinations of the
above tests in one strip, including:
• PT/INR Test Strips
• APTT Test Strips
• Coag Panel 2 Test Strips
1.3 Test Principle
qLabs
ElectroMeter Plus automatically detects the insertion of a qLabs
®
test strip and heats the strip to a preset operating temperature. After a
drop of blood is applied to the strip, the capillary channels carry the blood
to the reaction zones, where the blood mixes with pre-printed reagents
and starts to coagulate. Each strip contains two reaction zones. Each
reaction zone contains a pair of metallic electrodes, to which a constant
voltage is applied by the qLabs
of the blood proceeds, the current monitored across the two electrodes
changes. The qLabs
in the test zone and determines the PT and/or APTT results. The qLabs
ElectroMeter Plus and test strips provide APTT, PT and INR results.
1.4 Packaging
The qLabs
ElectroMeter Plus was packed individually.
®
Upon receipt of package, please open and remove the packaging
materials. Place the qLabs
connect the power adapter.
Do not keep the qLabs
heat source, or near an area with a strong magnetic field.
ElectroMeter Plus. As the coagulation
®
ElectroMeter Plus detects the change of the current
®
ElectroMeter Plus on a flat surface and
®
ElectroMeter Plus in direct sunlight, near a high
®
4
EN
®
®