ERROR CODE
E006
E007.x
E008
E009
E010
E011
E012
E013
E014
E015
DISCRIPTION
External liquid QC check out of
specifications.
PT calculation error.
Barcode scanner error.
INR result is out of range, or sample
may not be suitable for qLabs testing.
The operation of adding sample has
timed out.
You inserted a used test strip, or you
applied a blood sample to the test
strip before the meter warm-up was
complete.
APTT calculation error.
The strip is defective or improperly
operated when adding sample.
The value of HCT may be out of range
of 30%-55%.
Bluetooth initialization error.
33
CORRECTIVE
ACTIONS
1. Make sure the strip and liquid con-
trols are within the expiry date.
2. Make sure the Strip Code is con-
sistent with the code printed on
the liquid controls vial label.
3. Retest with a new strip.
Please retest with a new
test strip. If you receive a
second error code, then
contact your local distributor
and test the patient with a
laboratory method.
Restart the meter and retry.
Please retest with a new
test strip. If you receive a
second error code, then
contact your local distributor
and test the patient with a
laboratory method.
Please ensure the sample is added
within 2 minute after the 'add sample'
indicator is flashing
Retest with a new strip. Make sure to
add sample within 2 minutes after the
'add blood' indicator is flashing.
Please retest with a new
test strip. If you receive a
second error code, then
contact your local distributor
and test the patient with a
laboratory method.
Retest using a new strip. Make sure
the operation of adding sample is
exactly correct according to the user
manual guidance.
Patients with abnormal HCT value
may be recommended to test the INR
using another method.
Turn off the meter and restart the meter.
If the problem still exists, please contact
the technician for further assistant.
EN