NORMAL APTT
ON
NEXT
-
OK
Note:
APTT RATIO is used to evaluate the intrinsic coagulation function of a
patient (APTT RATIO=APTT Value/Normal APTT).
To get accurate results, NORMAL APTT shall be adjusted according to the
local clinical results. The NORMAL APTT is 36 by default. The default is
established by Micropoint with the test of 20 healthy patients.
8.6 Patient ID Setup
PATIENT ID
ON
BACK
-
OK
8.7 Operator ID Setting
OPERATOR ID
ON
BACK
-
OK
PASSWORD
---
NEXT
-
OK
Use "+" and "-" to change the highlighted
value. Press "NEXT" to switch between the
36
"ON" (or "OFF") and the value.
+
You can use "+" or "-" buttons to set the
input of Patient ID ON or OFF.
If you set the Patient ID ON, the meter will
prompt you to input the Patient ID before you
process a patient test.
+
Use "+" or "-" to turn on or turn off the
scanning and logout function of Operator ID.
When the status is at "ON" or "OFF", the
meter enters "PASSWORD" screen when
+
pressing "+" or "-".
Use "+" and "-" to change the highlighted
value. Press "NEXT" to switch between the
three digits. Press "OK" to confirm the entry.
The default password for the Operator ID is
123.
If the password is correctly entered, press
+
"OK" to return to the "OPERATOR ID" screen
and change the function switch from "OFF"
to "ON" or vice versa. Press "OK" to save the
settings.
15
EN
EN