5.3 Treatment value details –
standard models
You access the different programming
levels by repeated depressions of the
MENU key.
− You select the treatment parameters
with the corresponding parameter key.
− You change the treatment values with
the +/- keys and you enable/disable
functions by pressing the corresponding
parameter key again.
− You save the settings by pressing the
STOP key.
LEVEL 1:
n extension (stretching)
− maximum knee extension:
-10 degrees
− maximum hip extension:
7 degrees
n flexion (bending)
− maximum knee flexion:
120 degrees
− maximum hip flexion:
115 degrees
Note!
The programmed value and the value
measured at the patient's knee may
deviate slightly.
n speed
The speed can be adjusted between 5 %
and 100 % in steps of 5 %.
default setting: 50 %
n warm up protocol
During warm up, the patient will slowly
become used to the set maximum ex-
tension and flexion values, starting from
the middle position.
The warm up protocol starts in the
middle between the two maximum
ORMBS-12006_Artromot_K1_Comfort_2012_rz.indb 53
values set for stretching and bending.
The range of motion increases with each
cycle, until the programmed maximum
values are reached after a total of 15
cycles.
default setting: disabled
LEVEL 2:
n extension pause
Pauses occur at the extension limit, just
before the bending movement starts.
Pauses can be set to any value between
0 and 59 seconds in steps of 1 second,
and to values between 1 and 59 minutes
in steps of 1 minute.
default setting: no pause
n flexion pause
Pauses occur at the flexion limit, just
before the stretching movement starts.
Pauses can be set to any value between
0 and 59 seconds in steps of 1 second,
and to values between 1 and 59 minutes
in steps of 1 minute.
default setting: no pause
n therapy timer
Default setting is continuous operation
of the carriage. A clock symbol in the up-
per right-hand corner of the display iden-
tifies the continuous mode of operation.
The clock indicates the elapsed therapy
time.
In the continuous mode, the device
must be stopped with the STOP key.
However, you can also select therapy
durations of 1 to 59 minutes in steps of
1 minute and of 1 to 24 hours in steps of
30 minutes.
When the time has elapsed, the device
switches automatically off and stops in
the position:
extension + 10°. In this case, a circle
replaces the clock symbol. The circle fills
as the therapy time progresses.