9.2 Electromagnetic immunity
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
ARTROMOT®-K1 is intended for use in the electromagnetic environment specified below.
It is the responsibility of the customer or user to ensure that the ARTROMOT®-K1 unit is
used in such an environment.
Immunity test
Electrostatic
discharge (ESD) to
IEC 61 000-4-2
Electrical fast
transient/bursts
to
IEC 61 000-4-4
Surges to
IEC 61 000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines to
IEC 61 000-4-11
Power frequency
(50/60 Hz)
magnetic field to
IEC 61 000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
t
ORMBS-12006_Artromot_K1_Comfort_2012_rz.indb 65
IEC 60601
Compliance level
test level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV for power
± 2 kV for power
supply lines
supply lines
± 1 kV for input/
± 1 kV for input/
output lines
output lines
± 1 kV
± 1 kV
line-to-line
line-to-line
voltage
voltage
± 2 kV
± 2 kV
line-to-earth
line-to-earth
voltage
voltage
< 5 % U
< 5 % U
t
(> 95 % dip in U
)
(> 95 % dip in U
t
for ½ cycle
for ½ cycle
40 % U
40 % U
t
(60 % dip in U
)
(60 % dip in U
t
for 5 cycles
for 5 cycles
70 % U
70 % U
t
(30 % dip in U
)
(30 % dip in U
t
for 25 cycles
for 25 cycles
< 5 % U
< 5 % U
t
(> 95 % dip in U
)
(> 95 % dip in U
t
for 5 s
for 5 s
3 A/m
3 A/m
65
Electromagnetic
environment – guidance
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Mains power should be that of
a typical commercial or hospi-
tal environment.
Mains power should be that of
a typical commercial or hospi-
tal environment.
Mains power should be that of
t
)
a typical commercial or
t
hospital environment. If the
user of the ARTROMOT®-K1
t
unit requires continued
)
t
operation during power mains
interruptions, it is recommen-
ded that the ARTROMOT®-K1
t
)
unit be powered from an
t
uninterruptible power supply
or a battery.
t
)
t
Power frequency magnetic
fields should be at levels cha-
racteristics of a typical location
in a typical commercial or
hospital environment.