Maillefer Instruments
CH – 1338 Ballaigues
Switzerland
The handpiece is intended for use in the electromagnetic environment specified below. The Customer or the user of
the handpiece should assure that it is used in such an environment.
Immunity test
Electrostatic
±6 kV contact
discharge
±8 kV air
(ESD) IEC 61000-
4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1KV for input/
output lines
Surge
±1 kV differential
IEC 61000-4-5
mode
±2 kV common
mode
Voltage dips, short
<5 % UT for 0,5
interruptions and
cycle
voltage variations
(>95 % dip in UT )
on
power supply input
40 % UT for 5
lines
cycles
IEC 61000-4-11
(60 % dip in UT )
70 % UT or 25
cycles
(30 % dip in UT )
<5 % UT for 5 sec
(>95 % dip in UT )
The handpiece is intended for use in the electromagnetic environment specified below. The customer or the user of
the handpiece should assure that it is used in such an environment.
Immunity test
IEC 60601 test
Conducted RF
3Vrms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5
GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
ZF1902109.X / 03/ 2009 –updated 03/2011
IEC 60601
Compliance level
test level
±2, ±4, ±6KV
Contact
±2, ±4, ±8KV Air
Not applicable
Not applicable
Not applicable
Compliance
level
level
3 V
3 V/m
Electromagnetic environment –guidance
Floors should be wood, concrete
or ceramic tile. If floors are covered with
synthetic material, the relative humidity should
be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the handpiece requires continued operation
during power mains interruptions, it is
recommended that the handpiece be powered
from an uninterruptible power supply or a
battery.
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the handpiece, including
cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance
in metres (m).
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey,
compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with the following symbol:
a
should be less than the
b
6/36