Application; Indications For Use; Area Of Application; Conditions Of Use - Ottobock Kenevo 3C60 Instrucciones De Uso

Ocultar thumbs Ver también para Kenevo 3C60:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 123
In case of a system malfunction, safety mode makes restricted operation possible. Predefined
resistance parameters are configured in the product for this purpose (see Page 51).
The microprocessor-controlled hydraulic unit offers the following advantages
Stability while standing and walking
Smooth, harmonious, quiet initiation of the swing phase
Automatic recognition of sitting down. Manual unlocking of the joint not required.
Support while sitting down with individually adaptable resistance. This resistance remains
constant during the entire process of sitting down.
Support while standing up. The knee joint can be loaded even before reaching full extension.
Approximation of the physiological gait pattern
Adaptation of product characteristics to various surfaces, inclines, gait situations and walking
speeds
Manual locking of the knee joint for use of a wheelchair (see Page 50).

3 Application

3.1 Indications for use

The product is to be used solely for lower limb prosthetic fittings.

3.2 Area of application

Area of application according to the MOBIS mobility system:
This product is recommended for mobility grade 1 (indoor walker) and mobility
grade 2 (restricted outdoor walker). Approved for a body weight of max. 125 kg.
kg

3.3 Conditions of use

The product was developed for everyday use and should not be used for walking speeds over
3 km/h or unusual activities. These unusual activities include, for example, extreme sports (free
climbing, parachuting, paragliding, etc.).
Permissible ambient conditions are described in the technical data (see Page 58).
The prosthesis is intended for use exclusively on the user for whom the adjustment was made.
The manufacturer does not authorise use of the prosthesis on another person.

3.4 Indications

For users with knee disarticulation and transfemoral amputation
For unilateral or bilateral amputation
Dysmelia patients with residual limb characteristics corresponding to knee disarticulation or a
transfemoral amputation
The user must fulfil the physical and mental requirements for perceiving optical/acoustic sig­
nals and/or mechanical vibrations

3.5 Qualification

The product may be fitted only by qualified personnel authorised by Ottobock after completing the
corresponding training.

4 Safety

4.1 Explanation of warning symbols

Warning regarding possible serious risks of accident or injury.
WARNING
34

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Kenevo 3c60-st

Tabla de contenido