6.1 Connecting the power supply and battery charger

1
1) Slide the country-specific plug adapter onto the power supply until it locks into place (see
fig. 1).
2) Connect the round, three-pin plug of the power supply to the receptacle on the inductive
charger so that the plug locks into place. (see fig. 2)
INFORMATION: Ensure correct polarity (guide lug). Do not use force when connecting
the cable plug to the battery charger.
3) Plug the power supply unit into the outlet (see fig. 3).
→ The green LED on the back of the power supply lights up.
→ The yellow LED on the inductive charger lights up briefly to indicate the correct connec­
tion to the power supply.
→ If the green LED on the power supply does not light up and the yellow LED on the inductive
charger does not light up briefly while connecting the cable, there is an error (see Page 55).

6.2 Connect battery charger to the product

INFORMATION
Do not move the knee joint while it conducts the self-test immediately after disconnecting the
charger. Otherwise, an error may occur; if this happens, the problem can be corrected by
reconnecting and then disconnecting the charger.
INFORMATION
To make the operating time of the prosthesis as long as possible, the charger should not be
removed until immediately before the prosthesis is used.
42
2
1) Connect the inductive charger to the receiver of the charging unit
on the rear of the product. The charger is held in place by a mag­
net.
→ The correct connection of the battery charger to the product is
indicated by feedback (see Page 57).
2) The charging process starts.
→ Once the product battery is fully charged, the LED on the bat­
tery charger lights up green.
3) After the charging process is complete, hold the product still and
remove the inductive charger from the receiver.
→ A self-test is performed. The joint is operational only after cor­
responding feedback (see Page 57).