Guidance And Manufacturer's Declaration - Electromagnetic Immunity - DJO Global ARTROMOT Serie Modo De Empleo

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 136
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
ARTROMOT®-S3 is intended for use in the electromagnetic environment specified below. It
is the responsibility of the customer or user to ensure that the ARTROMOT®-S3 unit is used
in such an environment.
Immunity test
IEC 60601
test level
Conducted RF to
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF to
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cord-
less) telephones and land mobile radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic envi-
ronment due to fixed RF transmitters, an electromagnetic site survey should be con-
sidered. If the measured field strength in the location in which the ARTROMOT®-S3
exceeds the applicable ARTROMOT®-S3 unit should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be neces-
sary, such as reorienting or relocating the ARTROMOT®-S3 unit.
b) Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than
3 V/m.
Compliance
Electromagnetic
level
environment - guidance
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the
ARTROMOT®-S3 unit, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
3 Vrms
d = 1,2 √P
d = 1,2 √P at 80 MHz to 800 MHz
d = 2,3 √P at 800 MHz to 2,5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the trans-
mitter manufacturer and d is the
recommended separation distance
in meters (m).
3 V/m
Field strengths from fixed RF trans-
mitters, as determined by an electro-
magnetic site survey a
less than the compliance level in each
frequency rangeb
Interference may occur in the vici-
nity of equipment marked with the
follow ing symbol