Astralpool PC95 Manual De Instrucciones página 35

Ph – cloro – temperatura
Ocultar thumbs Ver también para PC95:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

GENERAL DESCRIPTION OF THE SIGNALLING LAMPS
PH measurement display
Chlorine measurement display
Two-coloured LED. This indicates the status of the pH pump
OFF
GREEN
RED
FLASHING RED
Two-coloured LED. This indicates the status of the chlorine pump
OFF
GREEN
RED
FLASHING RED
Red LED. This indicates an alarm status.
FLASHING RED
RED
Red LED. This indicates a status of total stoppage of the functions of the instrument.
RED
When it is illuminated, it signals a temporary block of operation of the pumps. The current outputs
remain locked on the last value that was re-transmitted.
CALIBRATION FUNCTIONS
7 pH
4 pH
Cl
2
The pH pump is switched off.
The set point is active however the pump is temporarily
switched off
The set point is active and the pump is switched on.
During programming, it indicates that the parameter being
displayed refers to control of the pH set point.
The chlorine pump is switched off.
The set point is active however the pump is temporarily
switched off
The set point is active and the pump is on.
In the programming phase, this indicates that the
parameter being displayed refers to control of the Chlorine
set-point.
An alarm status has occurred and the alarm relay is active.
To reset the alarm press the ACK key.
An alarm has been acknowledged by pressing the ACK
key, however the cause leading to the alarm is still
present. The alarm LED will extinguish once the cause of
the alarm has been eliminated.
The instrument prompts you to dip the electrode into a
standard solution having a pH of 7. It is therefore ready to
carry out calibration of the offset on the pH electrode
The instrument prompts you to dip the electrode into a
standard solution having a pH of 4. It is therefore ready to
carry out calibration of the gain on the pH electrode
The instrument is ready to carry out self-calibration of the
chlorine measurement. It is waiting for the chlorine value
measured by means of the DPD method to be entered.
6

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

136317

Tabla de contenido