Astralpool PC95 Manual De Instrucciones página 56

Ph – cloro – temperatura
Ocultar thumbs Ver también para PC95:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7.10___PROTECTING THE PROGRAMMING
The parameters that have been programmed can be protected by creating a bridge on the terminal
strip between terminals 12 and 14.
In this way it will no longer be possible to change the working data of the instrument.
If the
key is pressed, the indication DIS will appear on the display instead of PROG, to show
that the programming function is disable.
Protection of the programming is connected with step En , in which it is possible to enable the
functions for setting the SET and for calibration even when programming is protected.
EN
S- / -C / SC / --
Function for enabling the lower programming levels when protection of the programmed data is
provided on the terminal strip.
S-: the SET key remains active.
-C: the CAL key remains active.
SC: the SET and CAL keys both remain active.
--: No keys remain active.
NOTE: Terminals 12 and 14 can be connected to an external key-operated switch thanks to which
only authorised personnel will be in a position to enable the instrument.
7.11___RE-SETTING THE PARAMETERS
To eliminate the old calibrations and reset the default
parameters in the programming menu, proceed as follows:
Switch off the instrument
press the
key until the message RESP PAR
appears on the display.
Press the
key 10 seconds.
The instrument will show the message Save par.
27

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

136317

Tabla de contenido