6.1.2___ PRG2 menu
Enter the programming menu (See Page 16) and go to the PRG2 function.
The variables displayed are the following:
VARIABLES
PH SETTINGS
SET
ALL
ALH
REL
PRS
DEL
SPM
BP
REG
OUT
HIS
RMA
MMA
PHS
PHE
Ton
Tof
BND
PER
MON
DB
CHLORINE SETTINGS
SET
ALL
ALH
REL
PRS
DEL
SPM
BP
REG
OUT
HIS(*)
RMA
MMA
CLS
CLE
Ton
Tof
BND
PER
MON
DB
CTP
CPH
ENABLING
EN
ID
BR
PERMISSIBLE LIMITS
0 to 14.00
0 to SET
SET to 14.00
ACID/ALCA
NO / NC
0 to 255
20 to 150
0.50 to 2.80
ONOF/PULS/PROP
MIS/REG
0.02 to 1.40
0-20/ 4-20
0.1 to 5.00
0 to 10.00
PHS to 14.00
0 to 999 seconds
0 to 999 seconds
0.00 to 2.00 pH
0 to 999 seconds
0 to 255 seconds
0 to 10% BP
0 to 5.00
0 to SET
SET to 5.00
LO
NO / NC
0 to 255
20 to 150
0.50 to 2.80
ONOF/PULS/PROP
MIS/REG
0.02 to 0.50
0-20/4.20
0.1 to 5.00
0 to 4.00
PHS to 5.00
0 to 999 seconds
0 to 999 seconds
0.00 to 2.00
0 to 999 seconds
0 to 255 seconds
0 to 10% BP
0 to 10
ON / OFF
S-/-C/SC/--
1 to 30
1200 to 9600
DEFAULT VALUES
7.20 pH
6.50 pH
7.90 pH
ACID
NO
0 Seconds
20 strokes/min
0,50 pH
ONOF
MIS
0.00 pH
4-20 mA
0.1 mA
0 pH
14.00 pH
0 Seconds
0 Seconds
0.00 pH
500 Seconds
0 Seconds
0 % of BP
0,60 ppm
0,30 ppm
1,50 ppm
LO
NO
0 Seconds
20 Strokes/min.
0,50 ppm
ONOF
MIS
0.00 ppm
4-20 mA
0.1 mA
0 ppm
5.00 ppm
0 Seconds
0 Seconds
0.00 ppm
500 seconds
0 seconds
0 % of BP
0 %/°C
OFF
SC
1
2400
COMMENTS
SET POINT pH
pH minimum alarm
pH maximum alarm
Type of control
Status of the pH relay
Delay time, pH metering
Strokes/minute, Output
Freq.
Proportional band
Control mode
Working mode Output mA
Hysteresis
MA output dynamics
Offset Output mA
Initial value for reg. mA
End value for per reg. mA
ON time (pulsed control)
OFF time (pulsed control)
Safety range
Period (pulsed control)
Minimum ON time
dead band (reversed
hysteresis)
Chlorine SET POINT
Chlorine minimum alarm
Chlorine maximum alarm
Type of control
Status of chlorine relay
Delay time for chlorine
metering
Strokes/minute, Output
Freq.
Proportional band
Control mode
Working mode Output mA
Hysteresis
Output dynamics in mA
Output dynamics mA
Initial value for mA control
End value for per mA
control
ON time (pulsed control)
OFF time (pulsed control)
Safety band
Period (Reg. Prop.)
Minimum ON time
dead band (reversed hysteresis)
Coefficient for offsetting the the
measured value in relation to the
temperature expressed in %/°C.
Enabling of correction of the
chlorine measurement in relation to
the pH.
Enables the Cal and Set keys
Indicates the ID of the multi-point
.serial network
Baud rate of RS485 serial link