Immediately replace damaged power supply units, adapter plugs or battery chargers.
WARNING
Charging the prosthesis without taking it off
Risk of electric shock due to defects in the power supply unit or in the battery charger.
For safety reasons, remove the prosthesis prior to charging it.
CAUTION
Unauthorised manipulation of Axon-Bus components
Injury due to faulty control or malfunction of the Axon-Bus prosthetic system.
Manipulations to the Axon-Bus prosthetic system other than the tasks described in these
instructions for use are not permitted.
The battery may only be handled by Ottobock authorised, qualified personnel (no replace­
ment by the user).
The Axon-Bus prosthetic system and any damaged Axon-Bus components may only be
opened and repaired by authorised, qualified Ottobock personnel.
CAUTION
Signs of wear to system components
Injury due to faulty control or malfunction of the Axon-Bus prosthetic system.
In the interest of the user's own safety and in order to maintain the operational reliability and
the validity of the warranty, the specified service intervals must be observed.

4.4 Information on Proximity to Certain Areas

CAUTION
Proximity to sources of strong magnetic or electrical interference (e.g. theft prevention
systems, metal detectors)
Injury due to unexpected behaviour of the prosthesis system caused by interference with internal
data communication.
Avoid remaining in the vicinity of visible or concealed theft prevention systems at the
entrance/exit of stores, metal detectors/body scanners for persons (e.g. in airports) or other
sources of strong magnetic and electrical interference (e.g. high-voltage lines, transmitters,
transformer stations, computer tomographs, magnetic resonance tomographs, etc.).
When walking through theft prevention systems, body scanners or metal detectors, watch
for unexpected behaviour of the prosthetic system.
CAUTION
Distance to HF communication devices is too small (e.g. mobile phones, Bluetooth
devices, WiFi devices)
Injury due to unexpected behaviour of the prosthesis system caused by interference with internal
data communication.
Therefore, keeping the following minimum distances to these HF communication devices is
recommended:
• Mobile phone GSM 850/GSM 900: 0.50 m
• Mobile phone GSM 1800/GSM 1900/UMTS: 0.35 m
• DECT cordless phones incl. base station: 0.18 m
• WiFi (routers, access points,...): 0.11 m
• Bluetooth devices (third-party products not approved by Ottobock): 0.11 m