Information On The Power Supply/Battery Charging - Ottobock Axon-Bus Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 82
CAUTION
Risk of pinching between the hook tips
Injury due to pinching of body parts.
Ensure that no body parts are between the hook tips when using or rotating the product.
CAUTION
Gripping objects with different gripping force when using the "AxonHook" gripping
component
Injury due to unexpected behaviour of the gripping component.
Note that the gripping force can vary depending on the consistency (soft/hard) of the object
being gripped.
CAUTION
Use of the Axon-Bus prosthetic system with dirty or damaged contacts of the rotation
unit and gripping component
Injury due to unexpected behaviour of the Axon-Bus prosthetic system.
Make sure that the contacts of the rotation unit and gripping component are always clean
and free of grease.
Clean the electrical contacts regularly using cotton swabs and a mild soap solution.
Take care to avoid damaging the contacts with pointed or sharp objects.
NOTICE
Improper care of the Axon-Bus prosthetic system
Damage to the Axon-Bus prosthetic system due to the use of incorrect cleaning agents.
Clean the Axon-Bus prosthetic system with a damp cloth and mild soap only (e.g. 453H10=1
Ottobock DermaClean).
Use only the following products for cleaning/disinfecting the inner socket:
Cleaning: 453H10=1 Ottobock DermaClean
Disinfection: 453H16 Ottobock DermaDesinfect.

4.6 Information on the Power Supply/Battery Charging

CAUTION
Charging the Axon-Bus prosthetic system with soiled or damaged contacts
Injury due to unexpected behaviour of the Axon-Bus prosthetic system caused by insufficient
charging.
Make sure that the contacts are always clean and free of grease.
Clean the electrical contacts of the charging plug and charging receptacle regularly using
cotton swabs and a mild soap solution.
Take care to avoid damaging the contacts with pointed or sharp objects.
NOTICE
Charging the Axon-Bus prosthetic system with an incorrect power supply/battery char­
ger
Damage to the Axon-Bus prosthetic system due to incorrect voltage, current or polarity.
Use only power supply units/battery chargers approved for the Axon-Bus prosthetic system
by Ottobock (see instructions for use and catalogues).
28

Publicidad

Tabla de contenido
loading

Tabla de contenido