11 Pipetting (PIP)

Option
Add to Favorites
Learn pause time
Go to Multi-Disp
Operating Mode Help
11 Pipetting (PIP)
For information on the function of this mode, see Overview of modes, p. 75. For information on setting
the volume, speed, and other steps found in all modes, see Operation, p. 76.

11.1 Setting the volume

Prerequisite:
>
The operating mode has been selected.
a. Tap on the
b. Set the volume.

11.2 Filling a tip

11.3 Dispensing liquid and stopping liquid dispensing

a. Press the STEP button.
ð The tip is drained.
To stop dispensing liquid, tap on the X button.

11.4 Handling residual volumes

For information on handling residual volumes, see Switching to emptying, p. 80 and Dispensing liq-
uid, p. 80.
88
Operating manual
Meaning
Edits the time between the individual dispensing steps.
Starts the learn function.
Switches to the Multi-Disp operating mode. The volume set in Auto-
Disp mode is also set in the Multi-Disp operating mode.
Shows the help text for the mode.
button.
Prerequisite:
>
>
a. Hold the tip vertically in the vessel. During aspiration,
b. Press the STEP button.
ð The tip is filled until the set STEP volume or the nominal
To stop filling the tip, press the STEP button.
The volume is set.
On the touchscreen display, the message 'STEP to fill' ap-
pears.
make sure that the tip opening is always covered by liq-
uid, in order to prevent air bubbles in the tip.
volume of the tip is reached.
HandyStep® touch, HandyStep® touch S 997486