tation and as a result this can cause potential danger to the user.
Although biocompatibility tests have been performed on all the applied parts, under excep-
tional circumstances, allergic patients may have anaphylaxis. Do NOT use the monitor on
patient with anaphylaxis.
All connecting cables and rubber tubes of the applied parts should be kept away from the
patient's neck to prevent suffocation.
As a standard, please only use the components provided by the manufacturer or those that
are of the same model and specifications as the accessories.
If the monitor falls off a surface accidentally, please do NOT operate it before its safety and
technical performance have been tested, and positive results obtained.
Do NOT open the device cover without authorization. The cover should only be opened by a
qualified service personnel.
When disposing of the monitor and its accessories, the national regulation should be fol-
lowed.
There are some electromagnetic or inductance circuit designed in the device, use during MRI
environment could burns or adversely affect the MRI image or the device's accuracy. So the
device is MR unsafe.
The device and accessories are provided non-sterile.
The device has no alarm and is intended only for spot-checking.