The SpO
function has been adjusted before vending. If the user needs to adjust the SpO
2
adjust by using the simulator mode FLUKE INDEX2.
The adjustable units in the monitor such as the potentiometer cannot be adjusted without
permission so to avoid unnecessary failures that may affect the normal application.
It is recommended to use the battery once a month to ensure a strong power supply capacity
and long service life, and recharge once the power has completely run out..
8.1.3 Battery Maintenance
Please pay attention to the polarity of the battery, do NOT insert into the battery compart-
ment with reversed polarities.
In order to avoid damaging the battery, do NOT use other power supply device to charge the
battery.
After use, dispose of the battery according to local regulations, do NOT throw into fire.
Do NOT hit or strike the battery with force.
Do NOT use this battery in other devices.
Do NOT use this battery below -20°C or above 60°C.
In order to maintain the battery supply and prolong the battery lifetime, please charge the
battery routinely. Regularly, charge the battery every 3 months even if the device has not
been used.
Only use a battery with the specification recommended by the manufacturer.
Whether the monitor is on or off, the built-in battery will charge as long as the monitor is
connected to an AC adapter and the AC power is on. When the battery is full, it will stop
charging to avoid causing any damage. If the monitor is connected to an AC adapter and the
AC power is on, it will use the AC power, but when the AC power is off, the battery power will
be used. Priority of using the AC power and the power switch between the AC and battery is
ENGLISH