2.4 Blood Pressure Accuracy Check Method
Operation procedure:
1. Unscrew the M3x6 screw from the battery compartment on the back of the Spot-Check Mo-
nitor, as shown in figure 2.11.
2. Take a NIBP connector plug from the battery cover, as shown in figure 2.12. (Note: there are
two plugs but you will only need one.)
3. Air Path Connection: Take a piece of air tube (0.5~1m long, Φ8.0mm/Φ4.0mm diameter). Atta-
ch the NIBP connector with a connector plug on to one end of the air tube. Connect the other
end to the 3-way connector. Connect the other 2 ends of the 3-way connector to a Mercury
Sphygmomanometer as shown in Figure 2.13.