4.3 Pulse Rate Measurement

1. PR measuring range: 30bpm~240bpm
2. Pulse rate measuring accuracy: ±2bpm or ±2%, which is greater

4.4 Temperature Measurement

1. Measuring range: 32.0°C~43.0°C
2. Measuring accuracy: ±0.2°C is for TEMP range from 36.0°C to 39.0°C, and ±0.3°C is for the
rest; ±0.4°F is for TEMP range from 96.8°F to 102.2°F, and ±0.5°F is for the rest.
3. Response time: ≤5s

4.5 Blood Glucose Measurement (Optional)

1. Technique: Amperometric, glucose oxidase
2. Measuring range: 1.1mmol/L~33.3mmol/L (20~600mg/dL)
3. Measuring time: 6 seconds

4.6 ECG Measurement (Optional)

1. Heart Rate measuring range: 30bpm~240bpm
2. Heart Rate measuring accuracy: ±2bpm or ±2% whichever is greater
3. Display scale: 5.0mm/mV±10%
4. Common-mode rejection ratio (CMRR): ≥60dB