What if ...?
Error Description
Error Indication
Causes
Handpiece holder
Light meter does not respond
Blank display
Batteries inserted incorrectly
Batteries empty
Electronic defect
Handpiece
Indicator is
Handpiece overheated
Light cannot be activated
blinking
Indicator is
Power supply defective
blinking and
10 beeps are
Electronic defect
emitted
Blank display
Power supply defective
Electronic defect
10
Error rectification
Check batteries and insert them according
to the polarity indicated.
Insert new batteries according to the
polarity indicated.
Please contact your dealer or local
Service Center
Allow the handpiece to cool down and try
again after a while. If the error reappears,
please contact your dealer or local Service
Center.
Check your local power supply.
Please contact your dealer or local
Service Center.
Check your local power supply.
Please contact your dealer or local
Service Center.
Warranty
The warranty of bluephase C5 is 3 years from the date of
purchase. The warranty includes an LED operating life of
300 hours.
Malfunctions resulting from faulty material or
manufacturing errors are repaired free of charge during
the warranty period. The warranty does not provide the
right to recover any mateiral or non-material damage
other than those mentioned.
The apparatus must only be employed for the
intended use. Any other uses are contraindicated.
The manufacturer does not accept any liability resulting
from misuse and warranty claims cannot be accepted in
such cases. This is particularly valid for
– damage resulting from improper handling;
– damage of components resulting from wear under
standard operating conditions;
– damage resulting from external factors, e.g. impact
damage, dropping to the floor
– damage resulting from incorrect set-up or installation;
– damage resulting from connecting the unit to a power
supply whose voltage and frequency does not comply
with those stated on the rating plate;
– damage resulting from improper repairs or modifications
that have not been carried out by certified Service
Centers.
In case of a claim under warranty, the complete
apparatus (handpiece, handpiece holder, light probe)
must be sent C+F to the dealer or directly to Ivoclar
Vivadent, together with the purchase documents.
Use the original packaging with the corresponding
inserts for transportation.

Procedure in case of repair

Repair work may only be carried out by a certified Ivoclar
Vivadent Service Center. If a defect cannot be rectified,
please contact your dealer or local Service Center (see
addresses on the reverse side). A clear description of the
defect or the conditions under which the defect occurred
will facilitate locating the problem. Please enclose this
description when returning the apparatus.
Product specifications
Delivery form
1 Handpiece holder
3 Batteries (LR6 AA 1.5 VDC)
1 Handpiece with power cord and power pack
1 Light probe 10mm
3 Anti-glare cones
1 Set of Operating Instructions
Accessories
The following accessories are available for bluephase C5:
REF
Description
592 458
bluephase handpiece holder
592 475
Light probe 10 mm, black
551 752
Light probe 8 mm
592 482
Power-Booster light probe 13>8 mm, black
599 452
Pin-point light probe 6>2 mm, black
551 756
Anti-glare cones
592 496
Anti-glare shield
533 654
1 Heliotest
Technical data
Operating voltage
7.5 VDC
of handpiece
Power pack
100-240 VAC / 50-60HZ / max. 0.4A
Output 7.5 VDC / 1.7A
Operating conditions Temperature +10°C to +40°C
(50 to 104 °F)
Relative humidity 30% to 75%
Ambient pressure 700 hPa to 1060 hPa
Dimensions of
L = 200 mm, W = 120 mm
handpiece holder
H = 70 mm
Weight of handpiece 245 g
holder
Light source
5W LED
Wavelength range
430 – 490 nm
Light output
at minimum 500 mW/cm
2
Operation
6 min. on / 4 min. off (intermittent)
Dimensions of
L = 220 mm, W = 35 mm
handpiece
H = 160 mm
Weight of handpiece 200 g
Transportation and
Temperature –20°C to +70°C
storage conditions
(–4 to158 °F)
Relative humidity 10% to 75%
Ambient pressure 500hPa to 1060hPa
Store bluephase C5 in a closed, roofed room.
Protect the device from severe jarring.