9 Settings

FIG. 1
The Settings mode allows changing all parameters of each step.
It is accessed by long pressing the knob from the Operation mode
and leaved by also long pressing the knob or by running the
motor.
Note 1
All changes made in this mode are automatically saved for the
corresponding step.
Note 2
A. From the Settings mode menu, navigate through the operation
parameters by turning the knob CW or CCW.
The selected parameter symbol (1) is encased in a cyan square
and an arrow points on it.
B. If necessary, short press on the footpedal's orange button to go to
the next step without going back to the Operation mode.
The Settings mode is still displayed, the next step symbol
turns green and the step's last used settings are restored.
C. Short press on the knob to change the selected parameter
setting (setting sub-mode).
The selected setting sub-mode is displayed.

9.1 Operation mode

A. From the Settings mode menu, select the
short press on the knob to change operation mode.
Note 2
B. Turn the knob CW or CCW to alternatively toggle between
IMPLANTOLOGY
and SURGERY
C. Short press on the knob to exit operation mode setting.
Operation mode is saved and the Settings mode menu is
displayed again, FIG. 1.

9.2 MX-i LED micromotor speed

A. From the Settings mode menu, select the
short press on the knob to change maximum reachable speed.
FIG. 2
18
FIG. 1
symbol and
mode.
symbol and
1
B. Turn the knob CW or CCW to respectively increase or decrease
micromotor maximum reachable speed.
The speedometer (1) displays the set maximum reachable
speed.
C. Short press on the knob to exit speed setting.
New maximum reachable speed is saved and the Settings
mode menu is displayed again, FIG. 1.

9.3 MX-i LED micromotor torque

A. From the Settings mode menu, select the
short press on the knob to change maximum reachable torque.
FIG. 3
B. Turn the knob CW or CCW to respectively increase or decrease
micromotor maximum reachable torque.
The torquemeter (1) displays the set maximum reachable
torque.
C. Short press on the knob to exit torque setting.
New maximum reachable torque is saved and the Settings
mode menu is displayed again, FIG. 1.

9.4 MX-i LED micromotor rotation direction

A. From the Settings mode menu, select the
short press on the knob to change rotation direction.
Note 2
B. Turn the knob CW or CCW to alternatively toggle between
FORWARD
and REVERSE
C. Short press on the knob to exit rotation direction setting.
Rotation direction is saved and the Settings mode menu is
displayed again.
Note 3
FIG. 2
symbol and
symbol and
micromotor rotation.