11 List of errors & Troubleshooting

11.1 Safety warning (operating)

Warning description
Motor overheating
Release pedal
Low to high speed step tran-
sition
Motor jammed
Footpedal not connected
Motor not connected
22
Message
Speedometer blinking.
Cause of warning
Excessive power demand of
MX-i LED
the
micromotor.
• Pedal is pressed when
accessing settings sub-
modes.
• Pedal is pressed during device
startup.
• Pedal is pressed after
recovering from an error.
User switches from low speed
to high speed (≥ 100 RPM) step
in Implantology mode.
Motor is jammed for more than 2
seconds. Motor power supply is
cut to avoid overheating.
Pedal is not connected to
device.
Motor is not properly con-
nected to device,
Motor hardware is damaged.
Action
Avoid extended use. Let system cool down.
• Confirm setting by pressing the knob.
• Release footpedal and press it again.
• Release footpedal and press it again.
No action needed, the warning disappears after 2
seconds.
Release footpedal, release the bur and press
footpedal again.
Connect pedal to the device.
1. Acknowledge error.
2. (Re)connect the motor cable.
3. Press on the rotating knob.
4. If problem persists, contact Bien-Air Dental SA.