Appendix A: Emc Information; Guidance And Manufacturer's Declaration- Electromagnetic Emissions - Stryker SofCare Manual De Operación Y Mantenimiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 91

GUIDANCE AND MANUFACTURER'S DECLARATION- ELECTROMAGNETIC EMISSIONS:

This device is intended for use in the electromagnetic environment specified below. The user of this device should make
sure it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations / Flicker
emissions
IEC61000-3-3
WARNING
1. The device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, the device should be observed to verify normal operation in the configuration in which it will be used.
2.Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
3. Portable RF communications equipment (including peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part of the Pump, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
www.stryker.com

Appendix A: EMC Information

Compliance
Group1
Class B
Class A
Complies
476005-5210 V1.6
Electromagnetic Environment-Guidance
The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network
Return To Table of Contents
17

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

2890

Tabla de contenido