SV2 is intended for use in an electromagnetic environment specified below. The customer or the user of SV2 should
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations
Flicker Emissions
IEC 61000-3-3
Note: The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this
equipment might not offer adequate protection to radio-frequency communication services. The user might need to
take mitigation measures, such as relocating or re-orienting the equipment.
SV2 is suitable for use in the electromagnetic environment specified below. The customer or the user of SV2 should
Immunity Test
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrostatic fast
Transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
1-46

EMC information

Guidance and Manufacturer's Declaration - Electromagnetic Emissions
assure that it is used in such an environment.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
assure that it is used in such an environment.
IEC 60601 Test Level
+8 kV contact
+15 kV air
+2 kV for power supply
lines
+1 kV for input/output lines
+1 kV line(s) to line(s)
+2 kV line(s) to earth
Compliance
Group 1
Class A
establishments other than domestic
Class A
network that supplies buildings used
Complies
Compliance Level
+8 kV contact
+15 kV air
+2 kV for power supply
lines
+1 kV for input/output lines
+1 kV line(s) to line(s)
+2 kV line(s) to earth
KK-7500 REV 02
Electromagnetic Environment
SV2 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
SV2 is suitable for use in all
and those directly connected to the
public low voltage power supply
for domestic purposes.
Electromagnetic
Environment Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality is that
of a typical commercial
and/or hospital
environment.
www.stryker.com